fors_wave_calib - Derive dispersion relation from rectified arc lamp frame

SYNOPSIS

esorex fors_wave_calib [OPTIONS] FILE.sof

DESCRIPTION

This recipe is used to wavelength calibrate MOS/MXU slit spectra contained in the rectified arc lamp exposure produced with recipe fors_extract_slits.

A pattern-matching algorithm is applied as in recipe fors_detect_spectra.

The input spatial map is used in the production of the wavelength map.

Use recipe fors_wave_calib_lss for LSS data, or for MOS/MXU data where all slits have the same offset. For more details on this data reduction strategy please refer to the FORS Pipeline User´s Manual.

Note that specifying an input GRISM_TABLE will set some of the recipe configuration parameters to default values valid for a particular grism.

In the table below the MXU acronym can be alternatively read as MOS.

Input files:


DO category: Type: Explanation: Required:
SPATIAL_MAP_MXU Calib Spatial map Y
RECTIFIED_LAMP_MXU Calib Rectified arc exposure Y
SLIT_LOCATION_MXU Calib Slit location table Y
CURV_COEFF_MXU Calib Spectral curvature Y
MASTER_LINECAT Calib Line catalog Y
GRISM_TABLE Calib Grism table .

Output files:


DO category: Data type: Explanation:
REDUCED_LAMP_MXU FITS image Calibrated arc lamp exposure
DISP_COEFF_MXU FITS table Inverse dispersion coefficients
DISP_RESIDUALS_MXU FITS image Image of modeling residuals
WAVELENGTH_MAP_MXU FITS image Wavelengths mapped on CCD
SPECTRAL_RESOLUTION_MXU FITS table Spectral resolution table

OPTIONS

Expected spectral dispersion (Angstrom/pixel) (float; default: 0.0). The full name of this option for the EsoRex configuration file is fors.fors_wave_calib.dispersion [default = 0.0].
Initial peak detection threshold (ADU) (float; default: 0.0). The full name of this option for the EsoRex configuration file is fors.fors_wave_calib.peakdetection [default = 0.0].
Degree of wavelength calibration polynomial (int; default: 0). The full name of this option for the EsoRex configuration file is fors.fors_wave_calib.wdegree [default = 0].
Search radius if iterating pattern-matching with first-guess method (int; default: 4). The full name of this option for the EsoRex configuration file is fors.fors_wave_calib.wradius [default = 4].
Rejection threshold in dispersion relation fit (pixel) (float; default: 0.7). The full name of this option for the EsoRex configuration file is fors.fors_wave_calib.wreject [default = 0.7].
Name of line catalog table column with wavelengths (str; default: ´WLEN´). The full name of this option for the EsoRex configuration file is fors.fors_wave_calib.wcolumn [default = WLEN].
Start wavelength in spectral extraction (float; default: 0.0). The full name of this option for the EsoRex configuration file is fors.fors_wave_calib.startwavelength [default = 0.0].
End wavelength in spectral extraction (float; default: 0.0). The full name of this option for the EsoRex configuration file is fors.fors_wave_calib.endwavelength [default = 0.0].

Note that it is possible to create a configuration file containing these options, along with suitable default values. Please refer to the details provided by the 'esorex --help' command.

SEE ALSO

The full documentation for the fors pipeline can be downloaded as a PDF file using the following URL:

ftp://ftp.eso.org/pub/dfs/pipelines/fors/fors-pipeline-manual-4.9.pdf

An overview over the existing ESO pipelines can be found on the web page https://www.eso.org/sci/software/pipelines/.

Basic documentation about the EsoRex program can be found at the esorex (1) man page.

It is possible to call the pipelines from python using the python-cpl package. See https://packages.python.org/python-cpl/index.html for further information.

The other recipes of the fors pipeline are fors_align_sky(7), fors_align_sky_lss(7), fors_bias(7), fors_calib(7), fors_config(7), fors_dark(7), fors_detect_objects(7), fors_detect_spectra(7), fors_extract(7), fors_extract_objects(7), fors_extract_slits(7), fors_flatfield(7), fors_img_science(7), fors_img_screen_flat(7), fors_img_sky_flat(7), fors_normalise_flat(7), fors_photometry(7), fors_pmos_calib(7), fors_pmos_extract(7), fors_pmos_science(7), fors_remove_bias(7), fors_resample(7), fors_science(7), fors_spec_mflat(7), fors_subtract_sky(7), fors_subtract_sky_lss(7), fors_sumflux(7), fors_trace_flat(7), fors_wave_calib_lss(7), fors_zeropoint(7)

VERSION

fors_wave_calib 5.5.7

AUTHOR

Carlo Izzo <usd-help@eso.org>

BUG REPORTS

Please report any problems to usd-help@eso.org. Alternatively, you may send a report to the ESO User Support Department <usd-help@eso.org>.

LICENSE

This file is currently part of the FORS Instrument Pipeline Copyright (C) 2002-2010 European Southern Observatory

This program is free software; you can redistribute it and/or modify it under the terms of the GNU General Public License as published by the Free Software Foundation; either version 2 of the License, or (at your option) any later version.

This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. See the GNU General Public License for more details.

You should have received a copy of the GNU General Public License along with this program; if not, write to the Free Software Foundation, Inc., 51 Franklin St, Fifth Floor, Boston, MA 02110-1301 USA

5.5.7 fors_wave_calib